Home > Product > Level Meter

Level Meter

Contact us

TEL: 001-646-569-9958
FAX: 001-646-569-9958
Add:245 E MAIN ST STE 115,ALHAMBRA CA 91801,USA
> Read more

Level Meter

Omnigrad T TST310 RTD thermometer, cable probe

Omnigrad T TST310 RTD thermometer, cable probe

 

Cost efficient RTD sensor for use in many process and laboratory applications

 

Product Overview

 


The RTD cable probe is easy to install and provide a high operational safety due to reliable and accurate temperature measurement in common processes. The probe is specially suited to temperature measurement in machinery, laboratory equipment and plants in gaseous or liquid medium like air, water, oil, etc. Without additional thermowell the probe sheath is directly in contact with the process medium. This enables the cable probe to detect rapid temperature changes fast and accurate.

Benefits

  • High flexibility through user-specific insertion lengths and variable process connections

  • Fast response time

  • Single or double Pt100 sensor of accuracy class A, B, or AA as per IEC 60751

  • Types of protection for use in hazardous locations: Intrinsic safety (Ex ia) and non-sparking (Ex nA)

Field of application

The resistance thermometer is specially suited to temperature measurement in machinery, laboratory equipment and plants in gaseous or liquid media like air, water, oil and others.

Technical information

 

Measuring Principle Resistance Temperature Detector
Characteristic / Application metric style
cable probe
suitable for hazardous areas
process connection as compression fitting 
Thermowell / protection tube without (not intended to use with thermowell)
Insert / probe mineral insulated (MI), flexible
Outer diameter protection tube 3,0 mm (0,12'') 
6,0 mm (0,24'') 
Max. immersion length on request up to 10.000,00 mm (393,70'') 
Material protection tube/thermowell 1.4404 (316L)
1.4571 (316Ti)
Process connection compression fitting:
G1/4"
G1/2"
NPT1/4"
NPT1/2"
Tip shape straight 
Operating temperature range PT 100:
-50 °C ...400 °C
(-58 °F ...752 °F) 
Max. process pressure (static) at 20 °C: 25 bar (363 psi) 
Accuracy class A acc. to IEC 60751
class B acc. to IEC 60751
1/3 DIN class B acc. to IEC 60751
Response time t50 = 3 s
t90 = 6 s 
Integration head transmitter no
Ex - approvals ATEX II1/2D Ex ia IIIC,II1G Ex ia IIC T6
ATEX II1D Ex ia IIIC, II1G Ex ia IIC T6
ATEX II1G Ex ia IIC T6
ATEX II 3 G Ex nA IIC T6, II3D
NEPSI Ex ia IIC T6
IECEx Ex ia IIC T6 Ga
NEPSI Ex nA IIC T6

 

More Models and low price, please contact services@myjackwell.com